Smartphone-Based Colorimetric Microfluidic Paper-Based Analytical Device for On-Site Detection of Calcium Ions in Milk Samples

David Taiwo AjayiSiriwan TeepooDepartment of Chemistry,Faculty of Science and Technology,Rajamangala University of Technology Thanyaburi,Thanyaburi,Pathum Thani,Thailand
DOI: https://doi.org/10.1080/00032719.2024.2327638
2024-03-26
Analytical Letters
Abstract:Milk is a major dietary source of calcium; thus, accurate and precise quantification of calcium in milk and other dairy products is necessary for quality control. Conventional instrumental methods are sensitive and accurate but require expensive instrumentation, sample pretreatment, skilled personnel, and a long detection time. Herein, a microfluidic paper-based analytical device (μPAD) was developed to quantify calcium ions (Ca 2+ ) in milk using a smartphone as a detector. The method utilized a complexometric reaction between Ca 2+ and ethylene diamine tetraacetic acid with murexide as an indicator. The response was the average color intensity at the detection zones measured using the Color Picker application. The proposed μPAD exhibited a good linear detection range (1 to 10 mM), a low detection limit of 0.42 mM, and a short analysis time of 2 min. The determination of Ca 2+ in milk samples using the proposed μPAD was consistent with a potentiometric method. The proposed μPAD demonstrates many advantages, including rapid detection, acceptable accuracy, low cost, low sample volume, and suitability for on-site quantification of Ca 2+ in real samples.
chemistry, analytical
What problem does this paper attempt to address?