Reliability of Robotic Ultrasound Scanning for Scoliosis Assessment in Comparison with Manual Scanning

Maria Victorova,Heidi Hin Ting Lau,Timothy Tin-Yan Lee,David Navarro-Alarcon,Yongping Zheng
DOI: https://doi.org/10.48550/arXiv.2205.03566
IF: 3.7
2022-05-07
Robotics
Abstract:Background: Ultrasound (US) imaging for scoliosis assessment is challenging for a non-experienced operator. The robotic scanning was developed to follow a spinal curvature with deep learning and apply consistent forces to the patient' back. Methods: 23 scoliosis patients were scanned with US devices both, robotically and manually. Two human raters measured each subject's spinous process angles (SPA) on robotic and manual coronal images. Results: The robotic method showed high intra- (ICC > 0.85) and inter-rater (ICC > 0.77) reliabilities. Compared with the manual method, the robotic approach showed no significant difference (p < 0.05) when measuring coronal deformity angles. The MAD for intra-rater analysis lies within an acceptable range from 0 deg to 5 deg for a minimum of 86% and a maximum 97% of a total number of the measured angles. Conclusions: This study demonstrated that scoliosis deformity angles measured on ultrasound images obtained with robotic scanning are comparable to those obtained by manual scanning.