Calibration-free 3D ray-tracing beam hardening correction in computed tomography

Marco Seracini,Matteo Bettuzzi,Rosa Brancaccio,Maria Pia Morigi
DOI: https://doi.org/10.1016/j.nima.2024.169226
2024-03-12
Abstract:In this paper we introduce a calibration-free beam-hardening correction technique in computed tomography, which improves the results achieved by a state-of-the-art correction method. Two main practical points are here underlined: (1) the need to perform a beam hardening correction without previous calibration; (2) the need to consider the shape of the object in all its whole three-dimensional extent. The proposed method requires: (a) a first complete reconstruction of the object by means of a cone-beam filtered back projection algorithm; (b) a segmentation of the reconstructed volume; (c) the construction of the attenuation when varying path-length graph; (d) the interpolation of the experimental data on the graph with a convenient mathematical function; (e) the correction of the measured attenuation values; (f) to perform again the complete cone-beam reconstruction. The result is an effective beam-hardening calibration-free correction over a large variety of objects, easy to use and quite fast. The proposed method results particularly useful when both the measurement time and the control of the geometry of the system are limited, like the ones generally faced in Cultural Heritage diagnostics. Computed Tomography reconstructions for test samples and for real cases are provided to highlight the effectiveness and the advantages of the new suggested approach.