A patient-specific scatter artifacts correction method

Wei Zhao,Stephen Brunner,Kai Niu,Sebastian Schafer,Kevin Royalty,Guang-Hong Chen
DOI: https://doi.org/10.48550/arXiv.1501.04408
IF: 4.506
2015-01-19
Medical Physics
Abstract:This paper provides a fast and patient-specific scatter artifact correction method for cone-beam computed tomography (CBCT) used in image-guided interventional procedures. Due to increased irradiated volume of interest in CBCT imaging, scatter radiation has increased dramatically compared to 2D imaging, leading to a degradation of image quality. In this study, we propose a scatter artifact correction strategy using an analytical convolution-based model whose free parameters are estimated using a rough estimation of scatter profiles from the acquired cone-beam projections. It was evaluated using Monte Carlo simulations with both monochromatic and polychromatic X-ray sources. The results demonstrated that the proposed method significantly reduced the scatter-induced shading artifacts and recovered CT numbers.