Systematic effects on a Compton polarimeter at the focus of an X-ray mirror
M. Aoyagi,R. G. Bose,S. Chun,E. Gau,K. Hu,K. Ishiwata,N. K. Iyer,F. Kislat,M. Kiss,K. Klepper,H. Krawczynski,L. Lisalda,Y. Maeda,F. af Malmborg,H. Matsumoto,A. Miyamoto,T. Miyazawa,M. Pearce,B. F. Rauch,N. Rodriguez Cavero,S. Spooner,H. Takahashi,Y. Uchida,A. T. West,K. Wimalasena,M. Yoshimoto
DOI: https://doi.org/10.1016/j.astropartphys.2024.102944
2024-02-23
Abstract:XL-Calibur is a balloon-borne Compton polarimeter for X-rays in the
$\sim$15-80 keV range. Using an X-ray mirror with a 12 m focal length for
collecting photons onto a beryllium scattering rod surrounded by CZT detectors,
a minimum-detectable polarization as low as $\sim$3% is expected during a
24-hour on-target observation of a 1 Crab source at 45$^{\circ}$ elevation.
Systematic effects alter the reconstructed polarization as the mirror focal
spot moves across the beryllium scatterer, due to pointing offsets, mechanical
misalignment or deformation of the carbon-fiber truss supporting the mirror and
the polarimeter. Unaddressed, this can give rise to a spurious polarization
signal for an unpolarized flux, or a change in reconstructed polarization
fraction and angle for a polarized flux. Using bench-marked Monte-Carlo
simulations and an accurate mirror point-spread function characterized at
synchrotron beam-lines, systematic effects are quantified, and mitigation
strategies discussed. By recalculating the scattering site for a shifted beam,
systematic errors can be reduced from several tens of percent to the
few-percent level for any shift within the scattering element. The treatment of
these systematic effects will be important for any polarimetric instrument
where a focused X-ray beam is impinging on a scattering element surrounded by
counting detectors.
High Energy Astrophysical Phenomena,Instrumentation and Methods for Astrophysics