Preconcentrating extractive polymeric network in paper-based sensing of copper

Natalie Uhlikova,Ian McKelvie,Spas D. Kolev,Evelina Vassileva,Tony G. Spassov,M. Inês G.S. Almeida
DOI: https://doi.org/10.1016/j.microc.2024.110744
IF: 5.304
2024-05-27
Microchemical Journal
Abstract:The concept of using an extractive polymeric sensing network on a paper-based substrate for the pretreatment of samples in the colorimetric analysis of heavy metals is demonstrated with the determination of Cu 2+ in aqueous samples. The newly developed paper-based device combines the advantages of both liquid–liquid extraction-based separation and the analytical capabilities of microfluidic paper-based analytical devices (μPADs). The preconcentrating extractive polymeric network composed of 53.5 wt% poly(vinyl chloride) (PVC), 45 wt% extractant (di-(2-ethylhexyl)phosphoric acid (D2EHPA)) and 1.5 wt% dye (1-(2′-pyridylazo)-2-naphthol (PAN)), coated on filter paper, shows good water-wicking properties, allowing the transmission of high sample volumes. This results in significant analyte preconcentration. The extractant (D2EHPA) selectively binds Fe 3+ and Zn 2+ which otherwise will compete with Cu 2+ in its colour reaction with PAN at the chosen pH of 3.0. The device's capacity to process higher sample volumes resulting in higher sensitivity is increased by the integration of several filter paper absorption layers. The optimized device is characterized by a limit of detection (LOD) for tap and environmental waters of 3.4 and 9.6 μg/L, respectively; stability of 51 days if stored in a refrigerator; and total analysis time (from sample insertion to colour detection) of less than 5 min.
chemistry, analytical
What problem does this paper attempt to address?