Improved positional accuracy of dental implant placement using a haptic and machine‐vision‐controlled collaborative surgery robot: A pilot randomized controlled trial

Jun‐Yu Shi,Bei‐Lei Liu,Xin‐Yu Wu,Min Liu,Qi Zhang,Hong‐Chang Lai,Maurizio S. Tonetti
DOI: https://doi.org/10.1111/jcpe.13893
IF: 7.478
2023-10-24
Journal Of Clinical Periodontology
Abstract:Aim To compare the implant accuracy, safety and morbidity between robot‐assisted and freehand dental implant placement. Materials and Methods Subjects requiring single‐site dental implant placement were recruited. Patients were randomly allocated to freehand implant placement and robot‐assisted implant placement. Differences in positional accuracy of the implant, surgical morbidity and complications were assessed. The significance of intergroup differences was tested with an intention‐to‐treat analysis and a per‐protocol (PP) analysis (excluding one patient due to calibration error). Results Twenty patients (with a median age of 37, 13 female) were included. One subject assigned to the robotic arm was excluded from the PP analysis because of a large calibration error due to the dislodgement of the index. For robot‐assisted and freehand implant placement, with the PP analysis, the median (25th–75th percentile) platform global deviation, apex global deviation and angular deviation were 1.23 (0.9–1.4) mm/1.9 (1.2–2.3) mm (p = .03, the Mann–Whitney U‐test), 1.40 (1.1–1.6) mm/2.1 (1.7–3.9) mm (p
dentistry, oral surgery & medicine
What problem does this paper attempt to address?