A Strategy to Establish a Quality Assurance/Quality Control Plan for the Application of Biosensors for the Detection of E. coli in Water

Nikou Hesari,Nursel Kıratlı Yılmazçoban,Mohamad Elzein,Absar Alum,Morteza Abbaszadegan
DOI: https://doi.org/10.3390/bios7010003
2017-01-03
Biosensors
Abstract:Rapid bacterial detection using biosensors is a novel approach for microbiological testing applications. Validation of such methods is an obstacle in the adoption of new bio-sensing technologies for water testing. Therefore, establishing a quality assurance and quality control (QA/QC) plan is essential to demonstrate accuracy and reliability of the biosensor method for the detection of <i>E. coli</i> in drinking water samples. In this study, different reagents and assay conditions including temperatures, holding time, <i>E. coli</i> strains and concentrations, dissolving agents, salinity and pH effects, quality of substrates of various suppliers of 4-methylumbelliferyl glucuronide (MUG), and environmental water samples were included in the QA/QC plan and used in the assay optimization and documentation. Furthermore, the procedural QA/QC for the monitoring of drinking water samples was established to validate the performance of the biosensor platform for the detection of <i>E. coli</i> using a culture-based standard technique. Implementing the developed QA/QC plan, the same level of precision and accuracy was achieved using both the standard and the biosensor methods. The established procedural QA/QC for the biosensor will provide a reliable tool for a near real-time monitoring of <i>E. coli</i> in drinking water samples to both industry and regulatory authorities.