Research of a New Biosensor for Detecting Escherichia Coli O157:H7 with Electrochemical Impedance Spectroscopy
Deng ZHANG,Song-yue CHENG,Li-feng QING,Rong LI,Ping WANG,Yanbin LI
DOI: https://doi.org/10.3969/j.issn.1004-1699.2005.01.002
2005-01-01
Abstract:New biosensors based on electrochemical impedance spectroscopy (EIS) were developed for detection of Escherichia coli O157:H7 through the surface immobilization of monoclonal antibody onto indium tin oxide (ITO) electrode. The biosensor could detect the target bacteria with a detection limit of 4×10 3 CFU/mL. A linear response in the electron-transfer resistance for the concentration of E.coli cells was found between 4×10 3 and 4×10 6CFU/mL. The biosensor characterizes with high sensitivity, simple operation, enabling relatively rapid and reliable analysis. The result proved the prospect of EIS biosensors on application to biomedical and environmental detection of E.coli O157:H7.