A gold nanoparticle-based lateral flow immunoassay for atrazine point-of-care detection using a handhold scanning device as reader

Tieqiang Sun,Zehua Xu,Shuai Yuan,Xiao Liu,Zongfen Chen,Zhenyu Han,Wentao Liu,Longxing Fan,Han Yang,Zhiwei Qie,Baoan Ning
DOI: https://doi.org/10.1007/s00604-021-05146-9
IF: 5.7
2022-03-23
Microchimica Acta
Abstract:A method is described to achieve accurate quantitative detection of atrazine (ATZ) in maize by using lateral flow strips based on gold nanoparticles (GNPs) and a handheld scanning reader. GNPs of 15 nm in diameter were applied as label, and a lateral flow immune assay strip was prepared. The linear range was 5.01–95.86 ng mL−1 with a detection limit of 4.92 ng mL−1 in phosphate buffer, 4 times better than the readout by the naked eye. ATZ-spiked corn samples were also analysed. The accuracy of results of spiked samples was confirmed by ELISA and liquid chromatography-tandem mass spectrometry (HPLC), which proved the reliability of the proposed method. A handhold device with an optical scanning system was designed for on-site quantitative detection. Combined with the pretreatment, the assay could be completed in less than 20 min.
chemistry, analytical
What problem does this paper attempt to address?