E‐Suture: Mixed‐Conducting Suture for Medical Devices

Onni J. Rauhala,Liang Ma,Duncan J. Wisniewski,Shan Shao,Brandon Schumacher,Jose Ferrero Lopez,Mara Kaspers,Zifang Zhao,Jennifer N. Gelinas,Dion Khodagholy
DOI: https://doi.org/10.1002/adhm.202302613
IF: 10
2023-12-27
Advanced Healthcare Materials
Abstract:Abstract Modern implantable bioelectronics demand soft, biocompatible components that make robust, low‐impedance connections with the body and circuit elements. Concurrently, such technologies should demonstrate high efficiency, with the ability to interface between the body's ionic and external electronic charge carriers. Here, we present a mixed‐conducting suture, e‐suture, composed of silk, the conducting polymer poly(3,4‐ethylenedioxythiophene) polystyrene sulfonate (PEDOT:PSS) and insulating jacketing polymers. The resulting e‐suture has mixed‐conducting properties at the interface with biological tissue as well as effective insulation along its length. The e‐suture can be mechanically integrated into electronics, enabling acquisition of biopotentials such as electrocardiograms, electromyograms and local field potentials (LFP). Chronic, in vivo acquisition of LFP with e‐sutures remains stable for months with robust brain activity patterns. Furthermore, e‐sutures can establish electrophoretic‐based local drug delivery, potentially offering enhanced anatomical targeting and decreased side effects associated with systemic administration, while maintaining an electrically conducting interface for biopotential monitoring. E‐sutures expand on the conventional role of sutures and wires by providing a soft, biocompatible, and mechanically sound structure that in addition has multifunctional capacity for sensing, stimulation, and drug delivery. This article is protected by copyright. All rights reserved
nanoscience & nanotechnology,engineering, biomedical,materials science, biomaterials
What problem does this paper attempt to address?