Every minute counts: A comparison of thawing times and haemostatic quality of plasma thawed at 37°C and 45°C using four different methods

J. McCullagh,C. Booth,J. Lancut,S. Platton,P. Richards,L. Green
DOI: https://doi.org/10.1111/tme.13061
2024-06-24
Transfusion Medicine
Abstract:Background Having faster plasma thawing devices could be beneficial for transfusion services, as it may improve the rapid availability of thawed plasma for bleeding patients, and it might remove the need to have extended pre‐thawed plasma: thus, reducing unnecessary plasma wastage. Study Design and Methods The aims of this study were to assess (a) the thawing times and (b) in vitro haemostatic quality of thawed plasma using Barkey Plasmatherm V (PTV) at 37 and 45°C versus Barkey Plasmatherm Classic (PTC) at 37 and 45°C, Sarstedt Sahara‐III Maxitherm (SS‐III) at 37°C and Helmer Scientific Thermogenesis Thermoline (TT) at 37°C. Haemostatic quality was assessed using LG‐Octaplas at three different time points: baseline (5 min), 24 and 120 h after thawing. Results The thawing time (SD) of 2 and 4 units was significantly different between different thawers. PTV at 45°C was the fastest method for both 2 and 4 units (7.06 min [0.68], 9.6 min [0.87], respectively). SS‐III at 37°C being the slowest method (24.69 min [2.09] and 27.18 min [4.4], respectively) (p =