Method Comparison of Erythrocyte Sedimentation Rate Automated Systems, the VES-MATIC 5 (DIESSE) and Test 1 (ALIFAX), with the Reference Method in Routine Practice

Michele Cennamo,Loredana Giuliano,Gloria Arrigoni,Valentina Fardone,Roberta Russo,Luca Maria De Tomasi,Fabio Bertani,Gaetano Cammarota,Giovanni Brunetti,Lucia Del Vecchio,Michelarcangelo Partenope
DOI: https://doi.org/10.3390/jcm13030847
IF: 3.9
2024-02-02
Journal of Clinical Medicine
Abstract:Background: The erythrocyte sedimentation rate (ESR) is a routine and aspecific test that is still widely used. The reference-manual method for ESR determination is the Westergren method. The VES-MATIC 5 is a novel, fully automated, and closed system based on a modified Westergren method. This study conceived the aim of comparing two ESR analytical analysers, Test 1 and the VES-MATIC 5, with the reference method in routine practice. Methods: This study included 264 randomly analysed samples. A comparison between the two methods and Westergren was performed, and they were evaluated for inter-run and intra-run precision. In addition, we investigated possible interferences and different sensitivities to conventional analytes. Results: The comparison of methods by Passing–Bablok analysis provided a good agreement for both systems, with a better correlation for VES-MATIC 5 (p = 0.96) than Test 1 (p = 0.93), and sensitivity studies did not show any significant influence. Conclusions: The VES-MATIC 5 analyser demonstrated excellent comparability with the reference method, and it had better performance than Test 1. It can be employed in routine practice, bringing advantages such as a reduction in the probability of human error compared to the manual method, as well as an increase in operator safety and environmental protection.
medicine, general & internal
What problem does this paper attempt to address?