Photon-Counting Detector Computed Tomography (PCD-CT): A New Era for Cardiovascular Imaging? Current Status and Future Outlooks

Pietro G. Lacaita,Anna Luger,Felix Troger,Gerlig Widmann,Gudrun M. Feuchtner
DOI: https://doi.org/10.3390/jcdd11040127
2024-04-21
Journal of Cardiovascular Development and Disease
Abstract:Photon-counting detector computed tomography (PCD-CT) represents a revolutionary new generation of computed tomography (CT) for the imaging of patients with cardiovascular diseases. Since its commercial market introduction in 2021, numerous studies have identified advantages of this new technology in the field of cardiovascular imaging, including improved image quality due to an enhanced contrast-to-noise ratio, superior spatial resolution, reduced artifacts, and a reduced radiation dose. The aim of this narrative review was to discuss the current scientific literature, and to find answers to the question of whether PCD-CT has yet led to a true step-change and significant progress in cardiovascular imaging.
cardiac & cardiovascular systems
What problem does this paper attempt to address?
### Problems the Paper Attempts to Solve This paper aims to explore the potential application, current status, and future prospects of photon-counting detector computed tomography (PCD-CT) in cardiovascular imaging. Specifically, the paper focuses on the following aspects: 1. **Reducing Calcification Artifacts**: In traditional coronary CT angiography (CTA), high-density calcified plaques can cause "blooming" artifacts, leading to an overestimation of coronary artery stenosis. The paper investigates whether PCD-CT can reduce these artifacts by improving spatial resolution and using multi-energy imaging techniques, thereby enhancing the accuracy of coronary artery stenosis assessment. 2. **Improving the Accuracy of Coronary Stent Evaluation**: Traditional energy-integrating detector CT (EID-CT) has limitations in evaluating the patency of coronary stents, especially for smaller diameter or special types of stents. The paper assesses whether PCD-CT can more accurately evaluate coronary stent patency by reducing stent artifacts and improving image quality. 3. **Enhancing Soft Tissue Contrast and Reducing Radiation Dose**: PCD-CT converts X-ray photons directly into electrical signals, reducing electronic noise, enhancing soft tissue contrast, and potentially lowering radiation dose. The paper discusses the potential clinical value of these technical advantages. 4. **Application of Multi-Energy Imaging**: PCD-CT can perform multi-energy imaging, allowing the separation of specific substances based on their atomic energy thresholds. The paper explores the application of this capability in cardiovascular imaging, such as generating high-quality iodine maps or images of other materials (e.g., calcium or monosodium urate crystals). In summary, the goal of the paper is to evaluate whether PCD-CT can bring a real breakthrough in cardiovascular imaging, particularly in reducing calcification artifacts, improving the accuracy of coronary artery stenosis assessment, enhancing coronary stent evaluation, increasing soft tissue contrast, and reducing radiation dose.