Commissioning and first results of the 174 GHz collective Thomson scattering diagnostic at Wendelstein 7-X

D. Moseev,S. Ponomarenko,H.P. Laqua,T. Stange,S.K. Nielsen,H. Braune,G. Gantenbein,S. Illy,J. Jelonnek,W. Kasparek,A. Kuleshov,L. Krier,C. Lechte,S. Marsen,M. Nishiura,B. Plaum,R. Ragona,T. Ruess,M. Salewski,P. Stordiau,M. Thumm,R.C. Wolf,J. Zimmermann,the W7-X Team
DOI: https://doi.org/10.1088/1748-0221/19/03/c03056
2024-03-27
Journal of Instrumentation
Abstract:Collective Thomson Scattering (CTS) diagnostics measure the scattering spectrum of monochromatic incident radiation off collective fluctuations in the plasma. In this contribution, we present the first results from the upgraded CTS diagnostic at Wendelstein 7-X (W7-X) operating in the frequency range between 172 and 176 GHz. This frequency range allows for minimization of noise originating from the electron cyclotron emission in the plasma. Consequently, the good signal-to-noise ratio allows measurements of fast ions or bulk plasma parameters with higher temporal resolution compared with the previously used 140 GHz system.
instruments & instrumentation
What problem does this paper attempt to address?