Ultrasonic power and intensities produced by diagnostic ultrasound equipment

Paul L. Carson,Paula R. Fischella,Thomas V. Oughton
DOI: https://doi.org/10.1016/0301-5629(78)90076-5
1978-01-01
Abstract:The ultrasonic radiation force on an absorber was measured with a feedback microbalance to determine the ultrasonic power emitted into water by 26 diagnostic ultrasound instruments. Included in the measurements were pulse echo scanners, echocardiographic units, ophthalmic units, Doppler fetal heart detectors, fetal monitors, and continuous and pulsed Doppler units for peripheral vascular studies. The temporal and spatial average intensity at the transducer was calculated for those instruments by dividing the total power by the nominal area of the radiating surface. Estimates of the spatial maximum intensity in water, both temporal average and temporal peak, were made by measurements with miniature hydrophones normalized to the power measurements. Hydrophone measurements only were performed on linearly-scanned real time arrays. For transducers with single element transmitters, the maximum obtainable power ranged from 60μW to 37 mW, and maximum obtainable intensity at the transducer face ranged from 1 to 1700 W/m2. A much smaller range of power and intensity was encountered among systems designed for a given set of applications and providing similar performance in those applications. The temporal average and temporal peak intensities at the focal point of focused transducers employed frequently in the last few years for general purpose scanning were somewhat higher than previously quoted values because of the increased focusing.
radiology, nuclear medicine & medical imaging,acoustics