Content detection the endotoxin in rhGM-CSF by kinetic chromogenic method

Kong Wen-we
Abstract:Objective To establish the quality control method of quantitative testing of endotoxin in rhGM-CSF. Methods Kinetic chromogenic method was applied and quantitative method of testing endotoxin in rhGM-CSF was established via reliability experiment of standard curve and interference experiment. Results In the reliability experiment of standard curve,R2values of the curve were all above 0.98,which indicated the validity of the standard curve,in the interference experiment,the sample was redissolved and diluted five times before 0.5 EU/ml of standard endotoxin was added into the sample as recovered solution.The recovery rates of endotoxin were all close to 100%.Under the dilution ratio,the interference of sample concentration was minor to the endotoxin experiment,in addition,the contents of endotoxin in two batches of samples were all lower than the limited value through routine testing,and the test results were consistent with that of gel method. Conclusion The kinetic chromogenic method is applicable to the quantitative testing of endotoxin in rhGM-CSF.
Medicine,Chemistry