The effect of cooling rate on the precision of measurement of glow peak 5 in the thermoluminescence of LiF:Mg,Ti (TLD-100)

Yogev Bokobza,Shlomo Biderman,Yigal Shalom Horowitz,Leonid Oster,Ilan Eliyahu,Alex Shapiro,Dmitry Nemirovsky,Galina Reshes,Beny Herman,Hezy Einav
DOI: https://doi.org/10.1093/rpd/ncad318
2024-01-09
Radiation Protection Dosimetry
Abstract:Abstract The effect of natural rapid cooling and oven slow cooling on the precision of thermoluminescence measurements of LiF:Mg,Ti is investigated. Three separate series of measurements resulted in average precisions of 5.1 and 5.0%, respectively. However, the highest precision of 1.7% (1 SD) was achieved for an oven-cooled material.
environmental sciences,public, environmental & occupational health,radiology, nuclear medicine & medical imaging,nuclear science & technology