A new, simple, and accurate method for non-invasive estimation of pulmonary arterial pressure

J Xu,L-G Durand,P Pibarot
DOI: https://doi.org/10.1136/heart.88.1.76
IF: 5.7
2002-07-01
Heart
Abstract:OBJECTIVE: To develop and validate a new non-invasive method for the estimation of pulmonary arterial pressure (PAP) based on advanced signal processing of the second heart sound.DESIGN: Prospective comparative study.SETTING: Referral cardiology centre.PATIENTS: This method was first tested in 16 pigs with experimentally induced pulmonary hypertension and then in 23 patients undergoing pulmonary artery catheterisation.METHODS: The heart sounds were recorded at the surface of the thorax using a microphone connected to a personal computer. The splitting time interval between the aortic and the pulmonary components of the second heart sound was measured using a computer assisted spectral dechirping method and was normalised for heart rate.RESULTS: The systolic PAP varied between 14-73 mm Hg in pigs and between 20-70 mm Hg in patients. The normalised splitting interval was measurable in 97% of the recordings made in pigs and 91% of the recordings made in patients. There was a strong relation between the normalised splitting interval and the systolic PAP (pigs: r = 0.94, standard error of the estimate (SEE) = 5.3 mm Hg; patients: r = 0.84, SEE = 7.8 mm Hg) or the mean pulmonary pressure (pigs: r = 0.94, SEE = 4.1 mm Hg; patients: r = 0.85, SEE = 5.8 mm Hg).CONCLUSIONS: This study shows that this new non-invasive method based on advanced signal processing of the second heart sound provides an accurate estimation of the PAP.
cardiac & cardiovascular systems