A Multiplex "Disposable Photonics" Biosensor Platform and Its Application to Antibody Profiling in Upper Respiratory Disease

Michael R Bryan,Jordan N Butt,Zhong Ding,Natalya Tokranova,Nathaniel Cady,Brian Piorek,Carl Meinhart,Joshua Tice,Benjamin L Miller
DOI: https://doi.org/10.1021/acssensors.3c02225
IF: 8.9
2024-03-30
ACS Sensors
Abstract:Photonic technologies promise to deliver quantitative, multiplex, and inexpensive medical diagnostic platforms by leveraging the highly scalable processes developed for the fabrication of semiconductor microchips. However, in practice, the affordability of these platforms is limited by complex and expensive sample handling and optical alignment. We previously reported the development of a disposable photonic assay that incorporates inexpensive plastic micropillar microfluidic cards for sample...
chemistry, multidisciplinary, analytical,nanoscience & nanotechnology
What problem does this paper attempt to address?
The main problem that this paper attempts to solve is to develop a low - cost, disposable photonic biosensor platform capable of multiplexing for the detection of antibody profiles of upper respiratory tract diseases. Specifically, the research aims to overcome the limitations of existing singleplex photonic sensing systems in multiplexing capabilities, especially by simplifying the optical coupling and sample - handling processes to reduce cost and complexity. ### Key Issues 1. **Limitations of Existing Systems**: Existing "disposable photonic" sensor platforms can only perform singleplex detection because their optical configurations limit the reading of multiple waveguides. 2. **Requirement for Multiplexing**: In order to increase the density of diagnostic information and statistical significance, a multiplexing system that can detect multiple analytes simultaneously needs to be developed. 3. **Cost and Technical Challenges**: Multiplexing systems usually require complex fluid - handling subsystems, expensive consumables, and complex instrument read - out devices, which increase the complexity and cost of the system. ### Solutions To solve the above problems, the authors propose a new method to achieve multiplexed optical input/output (I/O) by mapping the output of each grating coupler on the photonic chip to a single multimode fiber in a fiber bundle. Specific improvements include: - **Fiber Bundle Design**: A hexagonally closely - packed fiber bundle is used to maximize the number of outputs within a given diameter and ensure a consistent core spacing. - **Free - Space Optics**: Each output grating is mapped to a single multimode fiber in the fiber bundle through free - space optics, simplifying the challenge of optical alignment. - **Compact and Low - Cost Design**: The new system can achieve multiplexed detection while maintaining low cost and simple operation. ### Application Examples The authors demonstrate the effectiveness of this system in detecting antibody responses against influenza and SARS - CoV - 2 antigens in human serum and saliva. This indicates that this platform is not only suitable for current public health needs (such as monitoring the immune status of COVID - 19 variants), but can also be extended to a wider range of clinically relevant tests. ### Summary This paper solves the problem of how to achieve multiplexed detection on a disposable photonic biosensor platform without significantly increasing complexity and cost. This progress is expected to promote the development of portable, low - cost medical diagnosis platforms, especially in resource - limited environments.