O. KhatirS. M. FekihA. SahliI. BoudjemaaA. BenkhettouM. SalemB. Bachir Bouiadjraa Department of Mechanical Engineering,Laboratory Mechanics Physics of Materials (LMPM),University of Sidi Bel Abbes,Sidi Bel Abbes,Algeriab Faculty of Mechanical Engineering,University of Sciences and Technology of Oran Mohamed Boudiaf (USTO-MB),El Mnaouer,Oran,Algeria
DOI: https://doi.org/10.1080/15376494.2024.2350677
2024-05-08
Mechanics of Advanced Materials and Structures
Abstract:Advancements in technology have revolutionized healthcare, with notable impacts on auditory health. This study introduces a novel approach aimed at optimizing materials for middle ear prostheses to enhance auditory performance. We developed a finite element (FE) model of the ear incorporating a pure titanium TORP prosthesis, validated against experimental data. Subsequently, we applied Functionally Graded Materials (FGM) methodology, utilizing linear, exponential, and logarithmic degradation functions to modify prosthesis materials. Biocompatible materials suitable for auditory prostheses, including Stainless Steel, titanium, and Hydroxyapatite, were investigated. Our findings indicate that combinations such as Stainless Steel with titanium and Hydroxyapatite offer improved outcomes compared to pure titanium and Hydroxyapatite ceramic, in terms of both displacement and stress. Additionally, personalized prostheses tailored to individual patient needs are feasible, underscoring the potential for further advancements in auditory healthcare.
materials science, multidisciplinary, composites,mechanics, characterization & testing