Design and Motion Control of Master–Slave Control Endotracheal Intubation Robot

Qian Qi,Junqi Lv,Ke Sun,Yi Sun,Andong Jiang,Aihong Ji
DOI: https://doi.org/10.1002/rcs.2669
2024-08-01
Abstract:ABSTRACT Background Master–slave remote control technology allows patients to be treated promptly during transport and also reduces the risk of contagious infections. Endotracheal intubation, guided by endoscopy and a master–slave system, enables doctors to perform the procedure efficiently and accurately. Methods In this paper, we present the development of a master–slave controlled endotracheal intubation robot (EIR). It is based on operation incremental mapping, a weighted recursive average filtering method to reduce vibration, and a virtual fixture designed to reduce mishandling in minimally invasive surgery. Results Simulation analysis of the master–slave control demonstrates that the weighted recursive average filtering method effectively reduces vibration, while the virtual fixture assists in confining the operator's movement within a delimited area. Experimental validation confirms the validity of the robot's structural design and control method. Conclusions The developed robot successfully achieves the necessary motion for endotracheal intubation surgery through master–slave control.
surgery
What problem does this paper attempt to address?