Design and Implementation of Master-Slave Control System for Nasal Continuum Minimally Invasive Surgery Robot

Tianliang Li,Qingsong Ai,Zaiwei Wang,Zemin Liu,Wei Meng
DOI: https://doi.org/10.23919/CCC63176.2024.10661402
2024-07-28
Abstract:To solve the problems encountered in transnasal surgery, such as instrument diameter coarseness, insufficient operation precision, restricted workspace, limited visual coverage, increased patient trauma, and heightened postoperative recurrence rates, this paper designs a robot operating system tailored for transnasal continuum minimally invasive surgery. The operation criteria for the nasal continuum minimally invasive surgical robot encompass the required reach and the bending angle within the nasal cavity and maxillary sinus lesion area. The initial step involves conducting kinematic analysis and modeling for the wire-driven flexible manipulator used in the surgical robot. Subsequently, a master-slave operating system for nasal minimally invasive surgery robot is devised with Touch master hand. The simulation results of the manipulator’s trajectory tracking confirm the effectiveness of its design. Additionally, the needle threading experiment validates the efficiency of the dominant-subordinate functioning structure in the minimally invasive nasal surgery robot. The continuous minimally invasive the surgical automaton suggested in this document is compact, offering a smaller footprint, and its operational space is optimized to fulfill surgical needs effectively.
Engineering,Medicine