A study of the possibility of fat quantification diagnosis technology in ultrasound image using machine learning

Seung-Man Yu
DOI: https://doi.org/10.1007/s40042-023-01004-4
2024-01-27
Journal of the Korean Physical Society
Abstract:In this study, after obtaining reference data of fat content by magnetic resonance imaging using a virtual phantom of fatty liver, an ultrasound image-based diagnostic technology that quantifies the fat content in the liver as a percentage using machine learning techniques through various variables in the ultrasound image. A virtual fatty liver phantom of oil/water 0–21% was fabricated so that water and fat could be homogeneously mixed. The ultrasonic frequency was set to 11.7 MHz and the dynamic range to 100, 150, 200 to obtain a homogeneous Dicom image according to the depth of each phantom, and post-processing was performed to show a homogeneous signal strength by filtering the noise. To get the standard data for quantifying fatty liver in the fatty liver phantom, data was acquired with a 3 T Dixon pulse sequence. An algorithm was developed using a machine learning method to develop a predictive model. When no filter was applied, the model with the highest coefficient of determination was the ensemble (boosting tree), and the coefficient of determination was 0.89. Moreover, the Root Mean Squared Error (RMSE) value was 2.5042. This study confirmed that an increase in fat content leads to ultrasound attenuation, and that intercept slope Gaussian regression could be used to quantify fat content with a high coefficient of determination and low RMSE.
physics, multidisciplinary
What problem does this paper attempt to address?