Development of the PSU Child Pump
Christopher Scheib,Raymond Newswanger,Joshua Cysyk,Karl Bohnenberger,Branka Lukic,Lichong Xu,Eric Yeager,Kirby Bletcher,Patrick Leibich,Quandashia Jackson,Heidi Flory,Mindy Tillinger,William Weiss,Gerson Rosenberg,Choon-Sik Jhun
DOI: https://doi.org/10.1097/mat.0000000000002202
2024-03-28
ASAIO Journal
Abstract:The Pennsylvania State University (PSU) Child Pump, a centrifugal continuous-flow ventricular assist device (cf-VAD), is being developed as a suitable long-term implantable device for pediatric heart failure patients between 10 and 35 kg, body surface area (BSA) of 0.5–1.2 m 2 , 1–11 years of age, and requiring a mean cardiac output of 1.0–3.5 L/min. In-vitro hydraulic and hemodynamic performances were evaluated on a custom mock circulatory loop with ovine blood. Normalized index of hemolysis (NIH) was evaluated under four conditions: 1) 8,300 rpm, 3.5 L/min, Δ P = 60 mm Hg, 2) 8,150 rpm, 5.1 L/min, Δ P = 20 mm Hg, 3) 8,400 rpm, 3.2 L/min, Δ P = 70 mm Hg, and 4) 9,850 rpm, 5.0 L/min, Δ P = 80 mm Hg, resulting in normalized index of hemolysis = 0.027 ± 0.013, 0.015 ± 0.006, 0.016 ± 0.008, and 0.026 ± 0.011 mg/dl, respectively. A mock fit study was conducted using a three-dimensional printed model of a 19 kg patient’s thoracic cavity to compare the size of the PSU Child Pump to the HeartMate3 and the HVAD. Results indicate the PSU Child Pump will be a safer, appropriately sized device capable of providing the given patient cohort proper support while minimizing the risks of blood trauma as they wait for a transplant.
engineering, biomedical,transplantation