Comparative steam distillation based digestion of complex inorganic copper concentrates samples followed by ion chromatographic determination of halogens

Nadeem Muhammad,Yun Zhang,Qamar Subhani,Azeem Intisar,Ye Mingli,Hairong Cui,Yan Zhu
DOI: https://doi.org/10.1016/j.microc.2020.105176
IF: 5.304
2020-11-01
Microchemical Journal
Abstract:<p>This study illustrates the feasibility of comparative steam distillation based digestion of standard copper concentrate samples for the effective extraction of fluoride (F<sup>−</sup>) and chloride (Cl<sup>−</sup>) contents and their subsequent ion chromatographic determination. Digestion of copper concentrate samples for the extraction of halides was carried out with a miniaturized and modified enclosed steam distillation unit. It is comprised of only a pair of Teflon tubes placed on an enclosed and automatic heating set up for respective sample digestion and steam generation. A multi-hole silicon chip was used for the cooling of the steam tube carrying volatile analytes, and at the end, a receiving bottle having 10 mL of 0.1 M NaOH was placed for the absorption of extracted halide contents from copper concentrate sample. Finally, the extracted fluorides and chloride content in the absorption bottle were injected in a pre-running ion chromatography system. Both analytes were isocratically separated by pumping automatically generated 20 mM KOH as a mobile phase through a Thermo Fisher Scientific anion exchange column IonPac® AS15 (250 mm × 4 mm i.d; 9 µm particle size) that further protected with a guard IonPac® AS15 (50 mm × 4 mm i.d; 9 µm particle size). The developed method showed good linearity for both halides in the range of 0.0020–5 µg/mL with an excellent correlation coefficient (0.99952). The limits of detection (LODs) of F<sup>−</sup> and Cl<sup>−</sup> were 0.0002 and 0.0005 µg/mL, respectively. Besides, the method showed excellent accuracy (87.95–100.02%) along with intra-and-inter-day precision of less than RSDs ≤ 1.41%. The developed method appeared robust and showed no matrix interference at analyte peaks of interest. Finally, this miniaturized and modified enclosed steam digestion unit compared with a conventional steam distillation unit. It appeared safer, reproducible, miniaturized (save reagents) and tap water free setup for the effective extraction of both halides from four complex inorganic copper concentrate samples.</p>
chemistry, analytical
What problem does this paper attempt to address?