Dissection-like appearance of focal catheter-induced vasospasm in intracoronary optical coherence tomography

Rick Volleberg,Peter Damman,Niels van Royen
DOI: https://doi.org/10.1093/eurheartj/ehae240
IF: 39.3
2024-04-27
European Heart Journal
Abstract:Intracoronary optical coherence tomography (OCT) enables high-resolution cross-sectional imaging of the coronary lumen and vascular wall. Accurate interpretation of OCT images is pivotal in the decision-making process and, among others, requires an adequate understanding of potential image artefacts. While expert consensus documents and state-of-the-art reviews have previously outlined common artefacts such as non-uniform rotational distortion and seam artefacts, 1,2 we present here an additional artefact that was not described in the respective documents but is often observed and might have significant clinical implications. Here, we present intracoronary OCT images of non-culprit lesions in five patients with an acute myocardial infarction. The images reveal a dissection-like appearance ( Figure 1 , left column). These features were predominantly located in non-diseased quadrants of the image and/or in non-diseased coronary segments (see Supplementary data online , Videos S1–S5 ). Angiographically, the OCT-detected image feature corresponded to a luminal indentation ( Figure 1 , middle columns). Repeat angiography after removal of the OCT imaging catheter and/or guidewire demonstrated complete resolution of the luminal indentation in all cases ( Figure 1 , right column).