Evaluation of tribological and biological properties of TaB 2 / PEEK composite coatings prepared by electrodeposition

Suyuan Huang,Lin Cao,Wei Li,Zhidan Lin,Peng Zhang
DOI: https://doi.org/10.1002/app.52265
IF: 3
2022-02-15
Journal of Applied Polymer Science
Abstract:Abstract It is well‐known that a lower wear rate, favorable mechanical properties, and limited inflammatory response are critical factors in determining the long‐term service of orthopedic implants. In this study, we prepare TaB 2 /PEEK coatings by electrophoretic deposition (EPD) of borated tantalum (TaB 2 ) and polyether‐ether‐ketone (PEEK) particles on a pure titanium plate (TA2). Nanoindenter, reciprocating ball‐on‐disc tester, and in vitro biological experiments were used to investigate the effects of TaB 2 particles on the mechanical properties, tribological properties, and biological properties of composite coatings. The results demonstrate that a small amount of TaB 2 particles were equally dispersed in the PEEK matrix, and the mechanical properties are improved, resulting in better tribological properties of the coating. In particular, the wear rate of P‐TB‐3 coating in the simulated body fluid (SBF) was 72% lower than that of the PEEK coating, with a friction coefficient (COF) of 0.164 and a wear rate of 1.45 × 10 −6  mm 3 ·N −1  m −1 , respectively. Furthermore, the P‐TB‐3 surface boosted Raw264.7 macrophage adherence and proliferation, with a potential to promote Raw264.7 macrophage transition to the M2 phenotype. These results suggest that the P‐TB‐3 coating prepared by EPD is considered as a promising implant replacement material.
polymer science
What problem does this paper attempt to address?