In Vivo Osseointegration Performance of Titanium Dioxide Coating Modified Polyetheretherketone Using Arc Ion Plating for Spinal Implant Application

Hsi-Kai Tsou,Meng-Hui Chi,Yi-Wen Hung,Chi-Jen Chung,Ju-Liang He
DOI: https://doi.org/10.1155/2015/328943
2015-01-01
BioMed Research International
Abstract:Polyetheretherketone (PEEK), which has biomechanical performance similar to that of human cancellous bone, is used widely as a spinal implant material. However, its bioinertness and hydrophobic surface properties result in poor osseointegration. This study applies a novel modification method, arc ion plating (AIP), that produces a highly osteoblast compatible titanium dioxide (TiO 2 ) coatings on a PEEK substrate. This PEEK with TiO 2 coating (TiO 2 /PEEK) was implanted into the femurs of New Zealand white male rabbits to evaluate its in vivo performance by the push-out test and histological observation. Analytical results show that AIP can prepare TiO 2 coatings on bullet-shaped PEEK substrates as implant materials. After prolonged implantation in rabbits, no signs of inflammation existed. Newly regenerated bone formed more prominently with the TiO 2 /PEEK implant by histological observation. The shear strength of the bone/implant interface increases as implantation period increases. Most importantly, bone bonding performance of the TiO 2 /PEEK implant was superior to that of bare PEEK. The rutile-TiO 2 coatings achieved better osseointegration than the anatase-TiO 2 coatings. Therefore, AIP-TiO 2 can serve as a novel surface modification method on PEEK for spinal interbody fusion cages.