The technical development of photon-counting detector CT

McCollough, Cynthia H.
DOI: https://doi.org/10.1007/s00330-023-09545-9
IF: 7.034
2023-04-05
European Radiology
Abstract:Since 1971 and Hounsfield's first CT system, clinical CT systems have used scintillating energy-integrating detectors (EIDs) that use a two-step detection process. First, the X-ray energy is converted into visible light, and second, the visible light is converted to electronic signals. An alternative, one-step, direct X-ray conversion process using energy-resolving, photon-counting detectors (PCDs) has been studied in detail and early clinical benefits reported using investigational PCD-CT systems. Subsequently, the first clinical PCD-CT system was commercially introduced in 2021. Relative to EIDs, PCDs offer better spatial resolution, higher contrast-to-noise ratio, elimination of electronic noise, improved dose efficiency, and routine multi-energy imaging. In this review article, we provide a technical introduction to the use of PCDs for CT imaging and describe their benefits, limitations, and potential technical improvements. We discuss different implementations of PCD-CT ranging from small-animal systems to whole-body clinical scanners and summarize the imaging benefits of PCDs reported using preclinical and clinical systems.
radiology, nuclear medicine & medical imaging