A High-Linearity Glucose Sensor Based on Silver-Doped Con A Hydrogel and Laser Direct Writing

Yulin Hu,Dasheng Yang,Hongbo Zhang,Yang Gao,Wenjun Zhang,Ruixue Yin
DOI: https://doi.org/10.3390/polym15061423
IF: 5
2023-03-14
Polymers
Abstract:A continuous glucose monitoring (CGM) system is an ideal monitoring system for the blood glucose control of diabetic patients. The development of flexible glucose sensors with good glucose-responsive ability and high linearity within a large detection range is still challenging in the field of continuous glucose detection. A silver-doped Concanavalin A (Con A)-based hydrogel sensor is proposed to address the above issues. The proposed flexible enzyme-free glucose sensor was prepared by combining Con-A-based glucose-responsive hydrogels with green-synthetic silver particles on laser direct-writing graphene electrodes. The experimental results showed that in a glucose concentration range of 0–30 mM, the proposed sensor is capable of measuring the glucose level in a repeatable and reversible manner, showing a sensitivity of 150.12 Ω/mM with high linearity of R2 = 0.97. Due to its high performance and simple manufacturing process, the proposed glucose sensor is excellent among existing enzyme-free glucose sensors. It has good potential in the development of CGM devices.
polymer science
What problem does this paper attempt to address?