Fully integrated flexible biosensor for wearable continuous glucose monitoring

Xiaofeng Jin,Guanhua Li,Tailin Xu,Lei Su,Dan Yan,Xueji Zhang
DOI: https://doi.org/10.1016/j.bios.2021.113760
IF: 12.545
2022-01-01
Biosensors and Bioelectronics
Abstract:Continuous physiological monitoring is a promising alternative to current chronic disease management for obtaining big data sets to help individualized therapy. Here, we present a continuous glucose monitoring platform consisting of a screen-printed electrochemical biosensor and a fully integrated wireless electrochemical analysis system. A biocompatible conjugated polymer (poly (N-phenylglycine)) was employed as the support material for enzyme immobilization. Specifically, a polyurethane outer layer was decorated onto the working electrode of the biosensor to construct a diffusion limiting membrane and improve the linear range of the glucose sensor. We optimized the fabricated glucose sensor so that it achieves a linear range of 1-30 mM and a sensitivity of 12.69 μA mM<sup>-1</sup>·cm<sup>-2</sup> in vitro. The long-term stability is up to 30 days by storing in PBS solution at 4°C. The overall system design was very small (0.8 × 1.8 cm) and consisted of a signal conditioning part, a programmable electrochemical chip, and a wireless connection using Bluetooth Low Energy with a smartphone. Finally, we carried out biocompatibility tests and animal experiments to demonstrate the device can successfully monitor blood glucose in vivo.
biotechnology & applied microbiology,biophysics,chemistry, analytical,electrochemistry,nanoscience & nanotechnology
What problem does this paper attempt to address?