Preparation and characterization of a homogenous immunoassay for point-of-care testing (POCT) of procalcitonin (PCT)

Zhaoying Li,Weixiang Zhai,Lu Wang,Jiyang Liu,Chunjie Li,Liang Xu
DOI: https://doi.org/10.1039/d3ay00890h
IF: 3.1
2023-08-31
Analytical Methods
Abstract:Procalcitonin (PCT) has been recognized as a specific and early marker for microbial infection and sepsis. Sensitive Measuring Interaction-triggered Luminescence Experiment (SMILE), a homogeneous immunoassay method, was established for point-of-care testing (POCT) of PCT. SMILE is achieved through the principle of double antibody sandwich, where two antibodies immobilized on the surface of polystyrene microspheres (donor and acceptor beads) bind to the PCT antigen, the donor bead contains phthalocyanine dye (luminol chemiluminescent substance), the acceptor bead contains dimethylthiophene derivatives and Eu chelates, therefore, singlet oxygen can be transferred when the distance between donor and acceptor beads is within 200 nm, and generating detectable luminescent signal. Scanning electron microscopy (SEM) was used to detect the diameter and Polymer dispersity index (PDI) of microspheres before and after binding with antibodies to characterize the immobilization of antibodies. The reaction conditions for antibody immobilization including pH, mass ratio and reaction time have also been optimized. The Limit of quantitation (LOQ) of SMILE method (0.01 ng/mL) was lower than that of LFI method (0.1 ng/mL), the working range (0.01-500 ng/mL) was wider than that of LFI method (0.1-50 ng/mL), and the assay time (10 min) was shorter than that of LFI method (15 min). So, SMILE is more suitable for POCT of PCT compared with lateral flow immunochromatography (LFI), which is the most used measuring method, due to its advantages of simple operation, timesaving, convenience, wide detection range, high sensitivity and accuracy.
spectroscopy,chemistry, analytical,food science & technology
What problem does this paper attempt to address?