Two annular CsI(Tl) detector arrays for the charged particle telescopes
Hong-Yu Zhu,Jian-Ling Lou,Yan-Lin Ye,Li-Sheng Yang,Bo-Long Xia,Kang Wei,Ying Chen,Wei-Liang Pu,Zi-Yao Hu,Hao-Yu Ge,Jia-Hao Chen,Kai Ma,Wen-Wu Wan,Hooi-Jin Ong,Yan-Yun Yang,Shi-Wei Xu,Kang Wang,Fang-Fang Duan,Jun-Bin Ma,Zhen Bai,Peng Ma,Ruo-Fu Chen
DOI: https://doi.org/10.1007/s41365-023-01319-8
2023-11-01
Nuclear Science and Techniques/Hewuli
Abstract:In this study, we constructed two annular detector arrays comprising 24 wedge-shaped CsI(Tl) crystals, and tested them using an source and radioactive beams of C on a CD target. We compared the properties of a CsI(Tl) crystal encapsulated with various reflectors, revealing that using the 80- m-thick ESR film to pack the CsI(Tl) crystal yielded the largest light output with the smallest non-uniformity in light output ( LO ). For the 24 CsI(Tl) detectors with the 80- m-thick ESR films, the average energy resolution improved as the average light output increased; however, it deteriorated as the LO value increased. To form two annular Si-CsI(Tl) telescopes for identifying the light-charged particles, the LO value and energy resolution of each CsI(Tl) detector were maintained under 20% and 7.7%, respectively. These telescopes were tested for the first time in a direct nuclear reaction experiment using C + d . The results demonstrated that the Z = 1 and Z = 2 charged particles were adequately discriminated by the telescopes using the standard - E method.