Rapid determination of 90 Sr in seawater using a novel porous crown-based resin and tandem quadrupole ICP-MS/MS in cool plasma and O 2 -He mode

Xingchen Liu,Yin Su,Feng Zhang,Jie Chen,Chenyang Peng,Kesheng Hu,Yiman Lu,Jing Lin,Shan Xing,Keliang Shi,Xiaolin Hou
DOI: https://doi.org/10.1016/j.sab.2024.107071
IF: 3.662
2024-11-06
Spectrochimica Acta Part B Atomic Spectroscopy
Abstract:There has been an increasing need for the rapid and automated analysis of 90 Sr in seawater for the purpose of radiological assessment and emergency response after the FDNPP accident. In this work, a rapid method for the determination of 90 Sr in seawater was established using ICP-MS/MS with cool plasma and collision-reaction gases combined with a new porous crown-based resin separation. A nanofiltration membrane was applied to rapidly pre-treat seawater samples within 3 min and remove most of the matrix components from 500 mL to 250 mL for the first time. A comprehensive investigation was performed to study the extraction behaviors of Sr, Zr, Ge, Y and matrix elements on the new resin, showing an excellent separation efficiency (> 10 3 ) using a U-shaped resin column. Based on their different first ionization potentials, the 90 Zr + and 89 Y + interferences were completely eliminated under cool plasma conditions. The interferences of 74 Ge 16 O + and 88 Sr were effectively suppressed by using 2 mL/min He-0.3 mL/min O 2 as the collision and reaction gases in the dynamic collision/reaction cell. Combined with chemical separation, the overall decontamination factors of Zr, GeO and Y were 1.9 × 10 10 , 6.9 × 10 7 and 6.9 × 10 7 respectively, and 88 Sr 1 H 2 + tailing was reduced by more than 200 times compared to a conventional ICP-MS. A detection limit of 4.6 pg/L (24.0 Bq/L) for 90 Sr was achieved with a sample analysis turn-around time of 6 h for a set of 12 samples. The method was validated by analyzing spiked seawater samples.
spectroscopy
What problem does this paper attempt to address?