A Rapid Method for Measuring Strontium-90 Activity in Crops in China

Lingjing Pan Pan,Guobing Yu,Deyun Wen,Zhi Chen,Liusi Sheng,Chung-King Liu,X. George Xu
DOI: https://doi.org/10.1051/epjconf/201715301003
2017-01-01
EPJ Web of Conferences
Abstract:A rapid method for measuring Sr-90 activity in crop ashes is presented. Liquid scintillation counting, combined with ion exchange columns 4', 4 "(5 ")-di-t-butylcyclohexane-18-crown-6,is used to determine the activity of Sr-90 in crops. The yields of chemical procedure are quantified using gravimetric analysis. The conventional method that uses ion-exchange resin with HDEHP could not completely remove all the bismuth when comparatively large lead and bismuth exist in the samples. This is overcome by the rapid method. The chemical yield of this method is about 60% and the MDA for Sr-90 is found to be 2.32 Bq/kg. The whole procedure together with using spectrum analysis to determine the activity only takes about one day, which is really a large improvement compared with the conventional method. A modified conventional method is also described here to verify the value of the rapid one. monitoring and emergency situation.
What problem does this paper attempt to address?