Advanced waveform analysis of diaphragm surface EMG allows for continuous non-invasive assessment of respiratory effort in critically ill patients at different PEEP levels

R. S. P. Warnaar,A. D. Cornet,A. Beishuizen,C. M. Moore,D. W. Donker,E. Oppersma
DOI: https://doi.org/10.1186/s13054-024-04978-0
IF: 15.1
2024-06-11
Critical Care
Abstract:Respiratory effort should be closely monitored in mechanically ventilated ICU patients to avoid both overassistance and underassistance. Surface electromyography of the diaphragm (sEMGdi) offers a continuous and non-invasive modality to assess respiratory effort based on neuromuscular coupling (NMCdi). The sEMGdi derived electrical activity of the diaphragm (sEAdi) is prone to distortion by crosstalk from other muscles including the heart, hindering its widespread use in clinical practice. We developed an advanced analysis as well as quality criteria for sEAdi waveforms and investigated the effects of clinically relevant levels of PEEP on non-invasive NMCdi.
critical care medicine
What problem does this paper attempt to address?
The paper attempts to address the issue of continuous, non-invasive monitoring of respiratory effort in mechanically ventilated patients in the Intensive Care Unit (ICU). Specifically, the authors aim to assess patients' respiratory effort using surface electromyography (sEMGdi) and investigate the impact of different levels of Positive End-Expiratory Pressure (PEEP) on the neuromuscular coupling index (NMCdi). ### Background and Problem 1. **Importance of Mechanical Ventilation**: Mechanical ventilation is used in the ICU to support the respiratory function of patients with respiratory failure. In assisted modes, the work of breathing is shared between the patient and the ventilator to reduce the patient's respiratory effort. 2. **Necessity of Monitoring Respiratory Effort**: Both excessive and insufficient assistance can lead to respiratory muscle dysfunction, necessitating close monitoring of respiratory effort to adjust the level of ventilator support. 3. **Limitations of Existing Monitoring Methods**: - Most mechanical ventilation devices allow for non-invasive but intermittent assessment of respiratory effort, such as through plateau pressure or end-expiratory occlusion pressure (Pocc). - Other methods like esophageal pressure (Pes) and invasive diaphragm electromyography (EAdi) are accurate but require specialized equipment and skills, and are time-consuming, making them unsuitable for routine clinical monitoring. 4. **Advantages and Challenges of Surface Electromyography (sEMGdi)**: - sEMGdi can provide continuous, non-invasive assessment of respiratory effort. - However, sEMGdi waveforms are prone to distortion due to interference from the heart and other muscles, affecting its application in clinical practice. ### Research Objectives 1. **Develop Advanced Signal Analysis Methods**: To improve the quality of sEMGdi waveforms, the authors developed a new automated advanced signal analysis method and set strict and lenient quality standards. 2. **Study the Impact of PEEP on NMCdi**: By conducting experiments at different PEEP levels, the authors aim to evaluate the impact of PEEP on non-invasive NMCdi. ### Methods 1. **Study Subjects**: 17 patients were recruited from a mixed ICU in a tertiary referral hospital in Enschede, Netherlands. 2. **Data Collection**: sEMGdi was measured using pre-gelled electrodes and active shielded electrode cables, while ECG signals and ventilator data were recorded simultaneously. 3. **Experimental Design**: Three spontaneous inspiratory efforts (Pocc) were performed at each PEEP level, and the corresponding sEMGdi waveforms were recorded. 4. **Signal Processing**: sEMGdi signals were band-pass filtered, QRS complexes were detected and cardiac interference was removed, and the sEMGdi envelope was calculated. 5. **Parameter Calculation**: Pocc and sEMGdi peaks were automatically detected, and the pressure-time product (PTPocc) and electrical activity-time product (ETPdi) were calculated. 6. **Data Analysis**: Generalized Estimating Equations (GEE) were used to analyze the impact of PEEP on NMCdi and to assess the impact of quality standards on data reproducibility and quality. ### Results 1. **Data Exclusion**: A large number of waveforms were excluded mainly based on low signal-to-noise ratio (SNR), irregular or abruptly stopped Pocc, and high area under the baseline (AUB). 2. **Reduction in NMCdi Variability**: After applying strict automated quality standards, the coefficient of variation of NMCdi decreased from 37% to 15%. 3. **Impact of PEEP on NMCdi**: With increasing PEEP, NMCdi significantly decreased, with NMCdi decreasing by 4.9 percentage points for every 1 cmH2O increase in PEEP. ### Conclusion 1. **Effectiveness of Advanced Signal Analysis**: The advanced signal analysis method can automatically identify high-quality sEMGdi waveforms, significantly reducing the variability of NMCdi. 2. **Impact of PEEP on NMCdi**: In mechanically ventilated ICU patients, NMCdi significantly decreases with increasing PEEP, consistent with findings in healthy subjects. 3. **Clinical Application Prospects**: These findings highlight the importance of considering PEEP-related diaphragm function when assessing sEMG-based respiratory effort, providing an important methodological basis for bedside continuous monitoring of respiratory effort.