Use of Carbon Dioxide-and Oxygen-TonometeredPhosphate- Bicarbonate-Chloride-Glycerol-Water Mixturesfor Calibration and Controlof pH, Pco2 and Po2ElectrodeSystems

A. H. Veefklnd,R. A. Van,A. Maas
Abstract:Calibration of pH, Pco2. and po2 electrode systems of modern blood-gas analyzers, designed with one sample cuvet for measurement, is mostly performed separately with buffer solutions and gas mixtures. We describe the development of solutions of known pH, Pco2 and Po2 for doing such calibrations simultaneously, containing phosphate, bicarbonate, and chloride in glycerol-water mixtures as solvent. A method is suggested for computing the relation between pH and log pco2 of these solutions in equilibrium with carbon dioxide gas. It is demonstrated that a solution of phosphate (Na2HPO4, KH2PO4, each 25 . mmol/liter), bicarbonate (NaHCO3, 30 mmol/ liter), and chloride (NaCI, 30 mmol/liter) in glycerolwater mixture (3/7 by vol) and equilibrated with CO2 In air (4 vol/100 vol) and CO2 in nitrogen (8 vol/lOG vol), respectively, makes possible accurate and simultaneous calibration of the pH, pco2, P02 electrodes of a Corning Model 165 blood-gasanalyzer. Similar solutions may also be used for quality-control of blood-gas measurement.