Real‐time imaging of decompression gas bubble growth in the spinal cord of live rats

Roman Alvarado,Ulrich M. Scheven,Jens‐Christian Meiners
DOI: https://doi.org/10.1002/mrm.30128
2024-04-24
Magnetic Resonance in Medicine
Abstract:Purpose To observe the growth and resolution of decompression gas bubbles in the spinal cord of live rats in real time using MRI. Methods We constructed an MRI‐compatible pressure chamber system to visualize gas bubble dynamics in deep tissues in real time. The system pressurizes and depressurizes rodents inside an MRI scanner and monitors their respiratory rate, heart rate, and body temperature while providing gaseous anesthesia under pressure during the experiments. Results We observed the formation of decompression gas bubbles in the spinal cord of rats after compression to 7.1 bar absolute and rapid decompression inside the MRI scanner while maintaining continuous gaseous anesthesia and vital monitoring. Conclusion We have shown the direct observation of decompression gas bubble formation in real time by MRI in live, anesthetized rats.
radiology, nuclear medicine & medical imaging
What problem does this paper attempt to address?