A-077 Evaluation of Diagnostic Thresholds for Adrenal Insufficiency Post-Cosyntropin Stimulation: Assessment Using Siemens Atellica Cortisol Immunoassay

S Li,L Chen,A T Calanche,J K Lee,L Ghalib,Q H Meng
DOI: https://doi.org/10.1093/clinchem/hvae106.076
IF: 12.114
2024-10-27
Clinical Chemistry
Abstract:BackgroundAccurate interpretation of the cosyntropin (adrenocorticotropic hormone [ACTH]) stimulation test relies on method- and assay-specific cortisol cutoffs. While the Endocrine Society recommends an 18 ug/dL cutoff, lower thresholds (14.6 ug/dL) were proposed for the Roche Elecsys II assay in 2021, which employs a monoclonal antibody. However, cutoff values for the new Siemens Atellica platform, utilizing polyclonal antibody-based cortisol assays, remain undetermined. This study aimed to evaluate cortisol thresholds specific to the Siemens Atellica immunoassay by comparing cortisol measurements with the Roche Elecsys II assay.MethodsThe ACTH stimulation test was conducted in patients with suspected adrenal insufficiency (n=28). Serum cortisol levels were measured using the Siemens Atellica and Roche Elecsys II assays. Atellica assay results were also compared with those obtained via liquid chromatography tandem mass spectrometry (LC-MS/MS).ResultsCortisol concentrations measured with the Atellica assay exhibited approximately 25-30% higher bias compared to LC-MS/MS and the Roche Elecsys II assay. Passing Bablok regression analysis revealed a strong correlation between the Atellica and LC-MS/MS assays (R 2 = 0.9312, slope = 1.26, intercept = 0.34). However, the correlation between Atellica and the Roche Elecsys II assay was weaker compared to LC-MS/MS (R 2 = 0.7049 at 0 min, slope = 1.28, intercept = 0.19; R 2 = 0.8066 at 30 min, slope = 1.28, intercept = 0.59; R 2 = 0.8584 at 60 min, slope = 1.17, intercept = 2.03). The optimized cortisol threshold using the Atellica assay was determined to be 18 ug/dL at 30 minutes after stimulation (sensitivity, 100%; specificity, 90%), using the proposed 14.6 ug/dL cutoff at 30 minutes on Roche Elecsys II as the standard.ConclusionsEmploying assay-specific cutoffs is crucial for reducing misclassification and overtreatment in patients with suspected adrenal insufficiency. We recommend a cortisol threshold of 18 ug/dL at 30 minutes after ACTH stimulation for the Atellica assay, consistent with the Endocrine Society's recommendation.
medical laboratory technology
What problem does this paper attempt to address?