Abstract:Objectives To assess the accuracy of autonomous robotic and fully guided static computer‐assisted implant surgery (sCAIS) performed on models and patients. Materials and Methods This study was divided into in vitro and in vivo sections. In vitro, 80 operators were assigned to two groups randomly. Forty operators performed forty autonomous robotic implant (ARI group) surgeries and the remaining forty operators carried out forty fully guided sCAIS (FGI group) surgeries on maxillary models, respectively. Each operator placed an implant in one maxillary model. In vivo, 60 patients with 113 implants from 2019 to 2023 (ARI group: 32 patients, 58 implants; FGI group: 28 patients, 55 implants) receiving implant surgeries were incorporated in this retrospective research. The preoperative and postoperative cone beam computer tomographs (CBCTs) were utilized to estimate the linear deviations and angular deviations in two‐dimensional (2D) and three‐dimensional (3D) space. The Pearson's chi‐square test, Shapiro–Wilk test, Student's t test, Mann–Whitney U test and mixed models were applied, and p <0.05 was considered statistically significant. Results In vitro, a total of 80 implants were enrolled and significant differences were found between the two groups (p
dentistry, oral surgery & medicine
What problem does this paper attempt to address?