An improved setup for radiocarbon analysis of water-soluble organic carbon in environmental matrices

Jan Strähl,Franziska A Lechleitner,Thomas Laemmel,Dylan Geissbühler,Gary A Salazar,Kaspar R Daellenbach,Sönke Szidat
DOI: https://doi.org/10.1017/rdc.2024.113
IF: 6.324
2024-12-08
Radiocarbon
Abstract:This paper presents an improved setup for radiocarbon analysis of water-soluble organic carbon based on wet chemical oxidation as installed at the Laboratory for the Analysis of Radiocarbon with AMS (LARA) at the University of Bern. The implementation of a non-dispersive infrared CO 2 detector allows more precise and accurate quantification of carbon amounts in samples and establishes the possibility of simple monitoring of the efficacy of flushing and sampling processes. A detailed blank assessment unveiled undesired oxidation of different materials and sample temperature as critical factors regarding the level of constant contamination. Contamination arising from oxidation of septum pieces and carbon-based glues in conventional sampling needles was minimized by developing a glass-sintered needle. This new needle was also designed to be longer, reducing the minimum amount of sample solution needed to 2 mL. The oxidation time and temperature (1 hr at 75°C) were optimized to further decrease contamination during analyses of samples with carbon amounts of up to ∼50 μg. With these improvements, we now report low constant contamination levels of 0.62 ± 0.12 μg C (with F 14 C of 0.19 ± 0.04), whereas the cross contamination factor was determined to be 0.25 ± 0.07%.
geochemistry & geophysics
What problem does this paper attempt to address?