Power deposition in the He bath induced by the Z gradient coil in the 11.7T Iseult magnet: theory versus measurements

Guy Aubert,Quentin GUIHARD,Olivier Dubois,Hervé Lannou,François-Paul Juster,Jean Belorgey,Francois Nunio,Alexis Amadon,Cédric Bonnelye,Vincent Gras,Caroline Le ster,Franck Mauconduit,Alexandre Vignaud,Lionel Quettier,Nicolas Boulant
DOI: https://doi.org/10.1088/1361-6668/ad4c12
2024-05-17
Superconductor Science and Technology
Abstract:Gradient-magnet interactions increase with higher field magnets and stronger gradient coils. Perhaps the most sensitive aspect is magnet quench, which can be induced by a loss of superconductivity of the main coil caused by a rise of the temperature of the He bath with gradient activity. Predicting power depositions thereby can be a very valuable tool to avoid dangerous frequency zones but also eventually correct design flaws. In this work, we report model predictions compared to measurements of power deposition in the He bath of the Iseult 11.7T magnet for the Z gradient coil axis.
physics, condensed matter, applied
What problem does this paper attempt to address?