The dilution evaluation as a corrective measure for interference in the white blood cell scattergram in Beckman Coulter DxH 900
Filipa P. Freitas,Jorge Reis,Joana Oliveira,Pedro Mota Veiga,Ana Raquel Paiva,Rui Soares
DOI: https://doi.org/10.1002/jcla.25007
IF: 3.124
2024-04-11
Journal of Clinical Laboratory Analysis
Abstract:Venous samples from oncologic patients were collected. All blood samples were processed in Beckman Coulter DxH 900 cell counter to elaborate a scattergram that differentiates the various white blood cells. Some samples presenting a population erroneously perceived as being lymphocytes. As a reference method and in order to evaluate if the poikilocytosis and lysis resistant to red blood cells are the cause of interference, it was done a differential leukocyte count. Another hypothesis for interference is hyperbilirubinaemia, and for that, blood samples were analysed in Atellica® CH 930 Analyser to measure total bilirubin and direct bilirubin. As a corrective measure, it was done a 1:3 dilution in all blood samples with interferences, being reprocessed in Beckman Coulter DxH 900. All the interferences were solved, and the main interference was Target Cells, which presented 97.4% of the cases, and hyperbilirubinaemia presented 74.3% of the cases. Background The Beckman Coulter DxH 900 is a haematological analyser capable of counting and sizing blood cells, and obtaining a complete blood cell count (CBC). This analyses different parameters of red blood cells (RBC), platelets and white blood cells/leukocytes. Some automated CBC counters present limitations due to specimen characteristics, abnormal cells or both factors. In the presence of abnormalities, the DxH 900 has a flagging system, warning the laboratory technician that something needs to be verified. In the present work, we evaluated samples from oncologic patients, presenting a population erroneously perceived as being lymphocytes. The most common explanations for this situation are RBC resistant to lysis or serum hyperbilirubinaemia. Methods In an attempt to solve and understand what the cause of this problem might be, we diluted our samples (1:3) and analysed the serum total bilirubin. To identify cells' abnormalities, the samples were also analysed by manual DLC counts. During the study, we also checked the different flags presented by the equipment. Results The results evidenced that the major interference was due to RBC lysis resistance, corresponding to 94.7% of the cases, while hyperbilirubinaemia was only present in 73.4%. Besides, we determined that some samples with normal bilirubin levels also presented interference, suggesting that hyperbilirubinaemia was not the main cause of the error. The most recurrent flag observed was "High event rate". Conclusion The dilution solved all of the observed interferences. The results between diluted and manual counts showed a strong correlation, leading us to introduce dilution in our laboratory routine.
medical laboratory technology