Preliminary evaluation for ultrasound‐guided targeted prostate biopsy using a portable surgical robot: Ex vivo results

Wenhe Jiang,Yongzhuo Gao,Mingwei Wen,Zhichao Ye,Huageng Liang,Dongmei Wu,Wei Dong
DOI: https://doi.org/10.1002/rcs.2597
2023-11-20
Abstract:Abstract Background Robotic systems are increasingly used to enhance clinical outcomes in prostate intervention. To evaluate the clinical value of the proposed portable robot, the robot‐assisted and robot‐targeted punctures were validated experimentally. Method The robot registration utilising the electromagnetic tracker achieves coordinate transformation from the ultrasound (US) image to the robot. Subsequently, Transrectal ultrasound (TRUS)‐guided phantom trials were conducted for robot‐assisted, free‐hand, and robot‐targeted punctures. Results The accuracy of robot registration was 0.95 mm, and the accuracy of robot‐assisted, free‐hand, and robot‐targeted punctures was 2.38 ± 0.64 mm, 3.11 ± 0.72 mm, and 3.29 ± 0.83 mm sequentially. Conclusion The registration method has been successfully applied to robot‐targeted puncture. Current results indicate that the accuracy of robot‐targeted puncture is slightly inferior to that of manual operations. Moreover, in manual operation, robot‐assisted puncture improves the accuracy of free‐hand puncture. Accuracy superior to 3.5 mm demonstrates the clinical applicability of both robot‐assisted and robot‐targeted punctures.
surgery
What problem does this paper attempt to address?