In Vivo Feasibility Study: Evaluating Autonomous Data-Driven Robotic Needle Trajectory Correction in MRI-Guided Transperineal Procedures

Mariana C Bernardes,Pedro Moreira,Dimitri Lezcano,Lori Foley,Kemal Tuncali,Clare Tempany,Jin Seob Kim,Nobuhiko Hata,Iulian Iordachita,Junichi Tokuda
DOI: https://doi.org/10.1109/lra.2024.3455940
Abstract:This study addresses the targeting challenges in MRI-guided transperineal needle placement for prostate cancer (PCa) diagnosis and treatment, a procedure where accuracy is crucial for effective outcomes. We introduce a parameter-agnostic trajectory correction approach incorporating a data-driven closed-loop strategy by radial displacement and an FBG-based shape sensing to enable autonomous needle steering. In an animal study designed to emulate clinical complexity and assess MRI compatibility through a PCa mock biopsy procedure, our approach demonstrated a significant improvement in targeting accuracy (p<0.05), with mean target error of only 2.2 ± 1.9 mm on first insertion attempts, without needle reinsertions. To the best of our knowledge, this work represents the first in vivo evaluation of robotic needle steering with FBG-sensor feedback, marking a significant step towards its clinical translation.
What problem does this paper attempt to address?