Measurement of BAS-TR imaging plate response to energetic aluminum ions

J. Won,J. Song,S. Palaniyappan,D. C. Gautier,W. Jeong,J. C. Fernández,W. Bang
DOI: https://doi.org/10.48550/arXiv.1912.10174
2019-12-19
Instrumentation and Detectors
Abstract:We measured the response of BAS-TR imaging plate (IP) to energetic aluminum ions in the 0 to 222 MeV energy range, and compared it with predictions from a Monte Carlo simulation code using two different IP models. Energetic aluminum ions were produced with an intense laser pulse, and the response was evaluated from cross-calibration between CR-39 track detector and IP energy spectrometer. For the first time, we obtained the response function of the BAS-TR IP for aluminum ions in the energy range from 0 to 222 MeV. Notably the IP sensitivity in the exponential model is nearly constant from 36 MeV to 160 MeV.
What problem does this paper attempt to address?