Piezoelectric Multi‐Channel Bilayer Transducer for Sensing and Filtering Ossicular Vibration

Muhammed Berat Yüksel,Ali Can Atik,Haluk Külah
DOI: https://doi.org/10.1002/advs.202308277
IF: 15.1
2024-02-22
Advanced Science
Abstract:This study presents a middle ear implantable transducer for fully implantable cochlear implants. The transducer`s dimensions are compatible with routine cochlear implant procedures. The multi‐channel system covers the daily acoustic band with a high signal‐to‐noise ratio. Objective audiological evaluations of speech signals underscore the encouraging results, affirming the feasibility of the middle ear implantable multi‐channel transducer for the first time. This paper presents an acoustic transducer for fully implantable cochlear implants (FICIs), which can be implanted on the hearing chain to detect and filter the ambient sound in eight frequency bands between 250 and 6000 Hz. The transducer dimensions are conventional surgery compatible. The structure is formed with 3 × 3 × 0.36 mm active space for each layer and 5.2 mg total active mass excluding packaging. Characterization of the transducer is carried on an artificial membrane whose vibration characteristic is similar to the umbo vibration. On the artificial membrane, piezoelectric transducer generates up to 320.3 mVpp under 100 dB sound pressure level (SPL) excitation and covers the audible acoustic frequency. The measured signal‐to‐noise‐ratio (SNR) of the channels is up to 84.2 dB. Sound quality of the transducer for fully implantable cochlear implant application is graded with an objective qualification method (PESQ) for the first time in the literature to the best of the knowledge, and scored 3.42/4.5.
materials science, multidisciplinary,nanoscience & nanotechnology,chemistry
What problem does this paper attempt to address?