Comparisons of an automated oscillometric device with a hybrid manual auscultatory device for the Korea National Health and Nutrition Examination Survey

Yu‐Mi Kim,Sang Hyeon Park,Jinho Shin,Ki‐Chul Sung,Sang‐Hyun Ihm,Seong Heon Kim,Dae‐Hee Kim,Sang Min Park,In Jeong Cho,Kyung Won Oh,Eun Mi Lee
DOI: https://doi.org/10.1111/jch.14805
2024-03-31
Journal of Clinical Hypertension
Abstract:This study evaluated an oscillometric device (OD), Microlife WatchBP Office AFIB, and a hybrid manual auscultatory device (AD), Greenlight 300TM, to determine a suitable blood pressure (BP) measurement device for the Korea National Health and Nutrition Examination Survey in a mercury‐free context. Adhering to the 2018 Universal Standard's suggested consensus, the study involved 800 subjects (mean age 51.2 ± 17.5 years; 44.3% male), who underwent triplicate BP measurements following 5 min of rest in a randomized order (OD‐first: 398 participants; AD‐first: 402 participants). BP difference was calculated as OD value minus AD value, with results stratified by measurement sequence. The overall BP difference and tolerable error probability were ‐1.1 ± 6.5/‐2.6 ± 4.9 mmHg and 89.2%/92.5% for systolic/diastolic BP (SBP/DBP), respectively. Lin's concordance correlation coefficient was 0.907/0.844 for SBP/DBP (OD‐first/AD‐first: 0.925/0.892 for SBP, 0.842/0.845 for DBP). The overall agreement for hypertension (BP ≥ 140 and/or 90 mmHg) was 0.71 (p