Interaction between left bundle branch area pacing lead and defibrillator lead: a case report.

Clara François,Benjamin De Becker,Maarten De Smet,Sébastien Knecht,Mattias Duytschaever,René Tavernier,Jean-Benoît le Polain de Waroux
DOI: https://doi.org/10.1016/j.hrcr.2023.10.026
2023-11-02
HeartRhythm Case Reports
Abstract:Key Teaching Points 1. Conduction system pacing by means of left bundle branch area pacing (LBBAP) is possibly a feasible approach as alternative for biventricular cardiac resynchronization therapy. 2 When an LBBAP lead and an ICD lead are combined, there is a risk of possible dangerous oversensing on the LBBAP-lead resulting in pacing inhibition, as well as oversensing on the shock lead, leading to an inappropriate shock. 3 We report a case of LBBAP lead interaction when combined with an implantable cardiac defibrillator (ICD) lead, due to tricuspid valve closure. Careful analysis of the EGM and the position of the leads in multiple views during implantation should be made when a shock lead is combined with a LBBAP lead.